HP (Hewlett-Packard) HP 16532A Sander User Manual


 
Table 1-3. Recommended Test Equipment
EQUIPMENT CRITICAL SPECIFICATIONS RECOMMENDED MODEL USE*
Signal Generator Frequency: 1 - 250 MHz at approx .
170 mV RMS resolution
Output Accuracy: ± 1 dB
1 MHz time base accuracy 0.25 ppm
HP 8656B Option 001 P
Dc Power Supply Range: 35.000 to + 35.000 Vdc, ± 1 mV HP 6114A P
Digital Multimeter 0.1 mV resolution
Accuracy: better than 0.005%
Resistance measurement: better than
0.25% accuracy
HP 3458A P
Power Meter/Sensor 1 - 250 MHz ± 3% accuracy HP 436/8482A P
Power Splitter Outputs differ by < 0.15 dB HP 11667B P
Blocking Capacitor 50 BNC (m-to-f), 0.18 µF, ± 200 V HP 10240B P
Adapter 50 BNC (m)(m) HP 1250-1236 C
Adapter 50 Type N(m)-to-BNC(m) HP 1250-0082 P
Adapter 50 BNC(f)-to-Dual Banana Plug HP 1251-2277 P
Adapter 50 Type N(m)-to-BNC(f) HP 1250-0780 P
Tee Adapter 50 BNC (m)(f)(f) HP 1250-0781 P,C
Cable Type N(m-to-m) 24-inch HP 11500B P
Cable 50 BNC (m-to-m) 48-inch HP 10503A P
Cable 50 BNC (m-to-m) 9 inch HP 10502A C
Cable 50 BNC-SMB HP 16532-61601 C
*P= Performance Tests C= Calibrations
HP 16532A - General Information
1-9