HP (Hewlett-Packard) HP 16532A Sander User Manual


 
3-9. VOLTAGE MEASUREMENT ACCURACY
This test verifies the DC voltage measurement accuracy of the instrument. This procedure uses a dual cur-
sor measurement that nullifies offset error, resulting in the following measurement:
± [(1.5% of full scale) + (0.016 x V/div)].
Specification:
± [(1.5% of full scale + offset accuracy) + (0.008 x V/div)]
Equipment Required:
The following equipment is required for this test. Procedures are based on the model or part numbers rec-
ommended.
Equipment Required Critical Specification
Recommended
Model/Part Number
DC Power Supply
Digital Multimeter
Cable
Adapter
(cable to power supply)
Adapter
Blocking Capacitor
-14 Vdc to + 14 Vdc, 0.1 mV resolution
Better than 0.1% accuracy
50 BNC (m to m) 48-inch
50 BNC (f) to Dual Banana Plug
50 BNC tee (m)(f)(f)
50 BNC (m to f) 0.18 µF, 200 V
HP 6114A
HP 3458A
HP 10503A
HP 1251-2277
HP 1250-0781
HP 10240B
HP 16532A - Performance Tests
3-6