Use the following table for steps 3 through 9.
V/Div Offset Supply Upper Limit Lower Limit
4 V/Div
1 V/Div
400 mV/Div
40 mV/Div
40 mV/Div
400 mV/Div
1 V/Div
4 V/Div
-7.0 V
-1.75 V
-700.0 mV
-70.0 mV
70.0 mV
700.0 mV
1.75 V
7.0 V
-14.0 V
-3.50 V
-1.40 V
-140.0 mV
140.0 mV
1.40 V
3.50 V
14.0 V
-13.7 V
-3.43 V
-1.37 V
-137.0 mV
143.0 mV
1.43 V
3.57 V
14.3 V
-14.3 V
-3.57 V
-1.43 V
-143.0 mV
137.0 mV
1.37 V
3.43 V
13.7 V
3. With the power supply disconnected from the channel input, set the Channel menu to the V/Div
range and offset values shown in the first line of the table.
4. On the oscilloscope, select RUN - REPETITIVE and wait for approximately five seconds, then select
STOP. Read the voltage displayed under the channel label on the left of the display. Enter this
value on the performance test record in the "Zero Input" column.
5. Set the power supply to the first supply value listed in the table. Connect the power supply to the
channel input as shown in figure 3-1.
6. On the oscilloscope, select RUN - REPETITIVE and wait for approximately five seconds, then select
STOP. Read the voltage displayed under the channel label on the left of the display. Enter this
value on the performance test record in the "Measured" column.
7. To obtain the test results, subtract the Zero Input voltage from the Measured voltage listed in the
performance test record. For example: (Measured) - (Zero Input) = (Actual). When calculating the
results, observe the plus or minus value of the voltage measurements. Record the results in the per-
formance test record in the "Actual" column.
8. Repeat steps 3 through 8 for the second line of the table, then for the rest of the lines of the table for
channel 1.
9. In the Waveform Selection menu, select Input channel 2. Delete channel 1 and insert channel 2. Re-
peat steps 2 through 8 for channel 2, replacing channel 1 with channel 2 where applicable.
HP 16532A - Performance Tests
3-8