Agilent Technologies E4446A Saw User Manual


 
Chapter 4 209
Instrument Functions: P - Z
Peak Search
Instrument Functions: P - Z
4.1 Peak Search
Places a marker on the highest peak and displays the search menu. If Peak Search Type
(Param) is set to Excursion & Threshold, the peak found must meet the defined peak
excursion and threshold values. (See
“Search Param” on page 215.) Peaks that are less than 1%
of the current span away from 0 Hz are ignored. For example, if Span is 1 MHz, peaks will
not be found between 10 kHz and +10 kHz. If no valid peak is found, an error (No Peak
Found) is displayed. To clear this message, press ESC before attempting another search.
The peak search parameters are Peak Threshold and Peak Excursion. All searches except
Peak Search (Next Peak, Next Left, Next Right, Peak Table, SCPI “PEAKS” command)
obey the Search Parameters, which means that only peaks which rise above the Peak
Threshold by at least the Peak Excursion are found. Peak Search obeys then when Peak
Search is in PARAM mode. When the Peak Search key is in MAX mode, the search
parameters are ignored for a Peak Search.
NOTE You can go into the Peak Search menu without actually performing a Peak
Search
by using the front-panel Return key (assuming you have previously
accessed the Peak Search menu).PressReturn to navigate through the
previously accessed menus until you return to the Peak Search menu.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:MAXimum
Remote Command Notes: The :CALC:MARK:PEAK:SEARC:MODE MAX|PAR
command specifies how a peak is identified for use with the marker
commands. See
“Peak Search” on page 218.
Example: CALC:MARK2:MAX performs a peak search using marker 2.
CALC:MARK2:Y? queries the marker amplitude (Y-axis) value for marker 2.
For more information on this command, see
“Marker” on page 177.
CALC:MARK2:X? queries the marker frequency or time (X-axis) value for
marker 2. For more information on this command, see
“Marker” on page 177.
4.1.1 Next Peak
Places the marker on the next highest peak with an amplitude less than the current peak.
The peak must meet the defined peak excursion and threshold values. Peaks that are less
than 1% of the current span away from 0 Hz are ignored. If no valid peak is found, an error
(No Peak Found)isdisplayed.PressESC to clear this message before attempting another
search. (Also see the
Peak Excursn and Pk Threshold key descriptions.)
Key Path:
Peak Search
State Saved: Not part of saved state.