Agilent Technologies E4446A Saw User Manual


 
416 Chapter5
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
Knob Increment: Steps through the available spans.
Step Key
Increment: Steps through the available spans.
Range: Span of your analyzer.
History: Added with firmware revision A.03.00
Remote Command:
This parameter can send up to 10 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:SPAN
[:SENSe]:HARMonics:RANGe[:LIST]:SPAN?
5.16.6.4 Res BW
Res BW is used to set the resolution bandwidth mode of the analyzer. If the range is zero
span and Res BW is set to Auto, the analyzer determines the optimum setting. Res BW set to
Man enables you to determine the setting.
Key Path:
Meas Setup, Range Table
Factory Preset: See Table 5-1 on page 414.
Knob Increment: Steps through the available resolution bandwidth filters.
Step Key
Increment: Steps through the available resolution bandwidth filters.
Range: Resolution bandwidth range of your analyzer.
History: Added with firmware revision A.03.00
Remote Command:
This parameter can send up to 10 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]:AUTO
OFF|ON|0|1
selects the mode.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]<integer>
The query for this parameter always returns 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:BWIDth|BANDwidth[:RESolution]:AUTO?