Agilent Technologies E4446A Saw User Manual


 
Chapter 5 357
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
Example: Sending fewer than six parameters to one of these commands will leave
the values of the unspecified offsets unchanged. If you don’t send settings
for all 6 offsets, it will set all the offsets that you specified, then it will set
any remaining offsets to the same setting as the last offset that you sent.
ACP:OFFS:LIST 50 Hz,75 Hz,100 Hz,125 Hz,150 Hz,175 Hz
ACP:OFFS:LIST:STAT ON,ON,ON,OFF,OFF,OFF
For example, after the above command is sent, sending the command with
only four parameters (ACP:OFFS:LIST:STAT ON,ON,ON,OFF) will result
in the fifth and sixth offset remaining the same as the previous setting
(OFF).
5.6.4.3 Ref BW
Sets the reference bandwidth (integration bandwidth) for each offset.
Key Path:
Meas Setup, Offset Setup
State Saved: Saved in instrument state.
Factory Preset: 2 MHz, if no radio standard (
None) is selected.
Default Terminator: Hz
Range: 300Hzto20MHz
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:INTegration]
<bw>,<bw>,<bw>,<bw>,<bw>,<bw>
[:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:INTegration]?
Example: ACP:OFFS:LIST:BAND 50,50,50,50,50,50
ACP:OFFS:LIST:BAND?
5.6.4.4 Pos Offset Limit
Enables you to set the upper limit for the upper segment of the specified offset pair.
Key Path:
Meas Setup, Offset Setup
State Saved: Saved in instrument state.
Factory Preset: 0.00 dB
Default Terminator: dB
Range: –200 dB to 200 dB