System
Instrument Functions: P - Z
4.8.5.1 GPIB Address
Shows the current GPIB address and allows you to change this value using the numeric
keyboard. The new value is displayed in the active function area. The GPIB port is always
active. The knob and step keys are not active for this function.
Key Path: System, Config I/O
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
SCPI Status Bits/
OPC Dependencies: None
Factory
Default: 18 (Reset by
System, Restore Sys Defaults.)
Terminators: Enter
Default Terminator: None
Resolution/Rounding/
Truncation: Integer
Knob Increment: 1
Step Key
Increment: 1
Range: 0 to 30
Remote Command:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <integer>
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?
Example: SYST:COMM:GPIB:ADDR 20
4.8.5.2 IP Address
Allows you to set the IP (internet protocol) address, domain name and node (host) name for
the instrument. The IP address of the instrument can be changed by entering a numeric
address composed of numbers and decimal points. Press
ENTER to complete the entry.
Key Path:
System, Config I/O
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
Factory
Default: 199.199.199.199 (Not reset by
System, Restore Sys Defaults.)
Remote Command:
:SYSTem:COMMunicate:LAN[:SELF]:IP <string>
:SYSTem:COMMunicate:LAN[:SELF]:IP?