Agilent Technologies E4446A Saw User Manual


 
410 Chapter5
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
Repeat—When you set Avg Mode to Repeat, after reaching the average count, all previous
result data is cleared and the average count is set back to 1.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: EXPonential
Range: EXPonential|REPeat
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:HARMonics:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:HARMonics:AVERage:TCONrol?
Example: HARM:AVG:TCON EXP
HARM:AVG:TCON?
5.16.3 Harmonics
Harmonics indicates the number of harmonics to measure before computing the total
harmonic distortion. The minimum number is 2 (only the fundamental and second
harmonic will be measured). The maximum number is 10.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10
Range: 2 to 10
Remote Command:
[:SENSe]:HARMonics:NUMBer <integer>
[:SENSe]:HARMonics:NUMBer?
Example: HARM:NUMB 5
5.16.4 ST/Harmonic
Sets the sweep time used to measure each harmonic. The value is set to 200 divided by the
resolution bandwidth, or 10 ms, whichever is greater when the measurement is started.
This sweep time is used only for measuring harmonics. The analyzer sweep time before the
measurement was started is used for finding the fundamental.
Key Path:
Meas Setup