Agilent Technologies E4446A Saw User Manual


 
310 Chapter5
One-Button Measurement Functions
MEASURE (Spectrum Analysis Mode)
One-Button Measurement Functions
Remote Command:
Measurement Results Available
Command n Results Returned
:CONFigure:MCPow
er
N/A Not Applicable
:FETCh:MCP[n]?
:MEASure:MCP[n]?
:READ:MCP[n]?
n=1 (or not
specified)
Returns 20 comma-separated scalar results, in the
following order.
1. First Carrier - relative power (dBc)
2. First Carrier - absolute power (dBm)
3. Second carrier frequency - relative power (dBc)
4. Second carrier frequency - absolute power (dBm)
5. lower offset A - relative power (dBc)
6. lower offset A - absolute power (dBm)
7. upper offset A - relative power (dBc)
8. upper offset A - absolute power (dBm)
9. lower offset B - relative power (dBc)
10. lower offset B - absolute power (dBm)
11. upper offset B - relative power (dBc)
12. upper offset B - absolute power (dBm)
13. lower offset C - relative power (dBc)
14. lower offset C - absolute power (dBm)
15. upper offset C - relative power (dBc)
16. upper offset C - absolute power (dBm)
17. NaN returned
18. NaN returned
19. NaN returned
20. NaN returned
If the results are not available, 999.0 is returned.
:FETCh:MCPower?
:MEASure:MCPower?
:READ:MCPower?
n=2...13
Returns absolute and relative values for carrier n-1. If results are not
available 999.0 is returned