Agilent Technologies E4446A Saw User Manual


 
Chapter 5 329
One-Button Measurement Functions
Meas Control
One-Button Measurement Functions
5.2 Meas Control
These functions allow you to pause and resume the currently selected measurement and to
select between continuous or single measurements.
NOTE If no measurement has been selected from the MEASURE menu, these
functions are not available.
Key Path: Front-panel key
5.2.1 Restart
This function restarts a previously paused measurement at the beginning. If the current
measurement is still in process, it will stop it as soon as possible and restart it from the
beginning.
Key Path: Front-panel key. It can also be found under
Meas Control.
Remote Command:
:INITiate:RESTart
Remote Command Notes: This command is equivalent to sending an :ABORt command
followed by an :INITiate[:IMMediate] command.
See “Abort the Sweep or
Measurement (Remote Command Only)” on page 331.
for more information.
Example: INIT:REST
5.2.2 Measure
Switches the analyzer between triggering the current measurement/sweep continuously or
triggering a single measurement. The front panel
Single key also puts the analyzer in
single-measurement mode.
Key Path:
Meas Control
State Saved: Save
Factory Preset: Continuous