Agilent Technologies E4446A Saw User Manual


 
Chapter 8 585
Using the STATus System
STATus Subsystem
Using the STATus System
Questionable Power Positive Transition
:STATus:QUEStionable:POWer:PTRansition <number>
:STATus:QUEStionable:POWer:PTRansition?>
This command determines which bits in the Questionable Power
Condition register will set the corresponding bit in the Questionable
Power Event register when the condition register bit has a positive
transition (0 to 1). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Temperature Register
Questionable Temperature Condition
:STATus:QUEStionable:TEMPerature:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Temperature Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Key Type: There is no equivalent front panel key.
Questionable Temperature Enable
:STATus:QUEStionable:TEMPerature:ENABle <number>
:STATus:QUEStionable:TEMPerature:ENABle?
This command determines which bits in the Questionable Temperature
Condition Register will set bits in the Questionable Temperature Event
register, which also sets the Temperature Summary bit (bit 4) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767