Agilent Technologies E4446A Saw User Manual


 
Chapter 2 61
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.0.10.1 Apply Corrections
Pressing
Apply Corrections (Yes) turns on the amplitude-correction factors. Corrections will
only be applied for the sets of correction factors whose correction state is set to On. To turn
a set of correction factors on, use the
Correction On Off key in the Antenna, Cable, Other, or
User menus.
Key Path:
AMPLITUDE / Y Scale, More, Corrections
Annunciation/
Annotation: When Apply Corrections (Yes) is selected, an A will appear in the screen
annotation on the left edge of the display, whether or not a correction set
has been turned on using the Correction (On) key in the Antenna, Cable,
Other, or User menus.
State Saved: Saved in Instrument State
Factory Preset: No
Remote Command:
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Remote Command Notes: To turn On or Off an individual correction set, use:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]
Example: CORR:CSET:ALL ON
CORR:CSET:ALL?
2.0.10.2 Antenna, Cable, Other, and User Keys
Pressing
Antenna, Cable, Other,orUser accesses the Correction menu for that type of
correction data. These 4 keys display the status of correction sets. If the key indicates On,
then amplitude corrections for this type have been enabled. To perform the corrections,
both this key and
Apply Corrections must indicate Yes. The status is toggled in the
correction set menu located under Correction.
Key Path:
AMPLITUDE / Y Scale, More, Corrections
Remote Command:
There is no remote command for this key.