Agilent Technologies E4446A Saw User Manual


 
Chapter 5 423
One-Button Measurement Functions
Meas Setup (Intermod (TOI))
One-Button Measurement Functions
5.18 Meas Setup (Intermod (TOI))
When Intermod (TOI) measurement has been selected in the Measure menu of the Spectrum
Analysis Mode, this key displays the appropriate measurement setup menu for third order
intercept (TOI).
The TOI measurement computes and displays the output intercept point (IP3), and places
markers upon the trace to indicate the measured signals and third-order products.
Key Path: Front-panel key
Dependencies/
Couplings: Menu changes depending on the Mode and Measurement selected.
History: Added with PSA firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.18.1 Avg Number
Set Avg Number to On to specify the number of measurement averages used when
calculating the measurement result. The average will be displayed at the end of each
sweep. Setting
Avg Number to Off disables measurement averaging.
Key Path:
Meas setup
State Saved: Saved in instrument state.
Factory Preset: 10 averages / Off
Range: 1 through 1000
History: Added with PSA firmware revision A.02.00
Remote Command:
[:SENSe]:TOIN:AVERage:COUNt <integer>
[:SENSe]:TOIN:AVERage:COUNt?
[:SENSe]:TOIN:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TOIN:AVERage[:STATe]?
Example: TOIN:AVER ON
TOIN:AVER?