Agilent Technologies E4446A Saw User Manual


 
Chapter 5 381
One-Button Measurement Functions
Meas Setup (Complementary Cumulative Distribution Function—CCDF)
One-Button Measurement Functions
5.10 Meas Setup
(Complementary Cumulative Distribution Function—CCDF)
When Power Stat CCDF has been selected in the Measure menu of the Spectrum Analysis
mode, this key displays the appropriate measurement setup menu for the complementary
cumulative distribution function (CCDF).
Power Complementary cumulative distribution function (CCDF) curves characterize the
higher-level power of the signal. It provides the distribution of peak-to-average power
ratios versus probability. A CCDF curve is defined by how much time the waveform spends
at or above the specified power level. The percent of time the signal spends at or above this
level defines the probability for that particular power level.
Key Path: Front-panel key
History: Added with firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.10.1 Meas BW
Press Meas BW to set the measurement 3 dB bandwidth. The range is 10 kHz to 8 MHz,
using the available RBW filters.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: Defined by the radio standard selected.
Terminators: Hz, kHz, MHz, GHz
Knob Increment: Steps through available Res BW filters.
Step Key
Increment: Steps through available Res BW filters.
Range: Lowest non-zero Span supported by Analyzer to the maximum Span of the
ESA
History: Added with firmware revision A.02.00
Remote Command:
[:SENSe]:PSTatistic:BANDwidth|BWIDth <freq>
[:SENSe]:PSTatistic:BANDwidth|BWIDth?
Remote Command Notes: The entered value is rounded to the nearest Res BW filter
frequency.