Agilent Technologies E4446A Saw User Manual


 
440 Chapter5
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
FactoryPreset: 0dB,0dB,0dB
Default Terminator: dB
Knob Increment: 1
Step Key
Increment: 5
Range: –200 dB to 200 dB
History: Added with firmware revision A.03.00
Remote Command:
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA <dB>,<dB>,<dB>
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA?
Example: CALC:MCP:OFFS:LIST:LIM:POS:DATA 50dB,55dB,60dB
:CALC:MCP:OFFS:LIST:LIM:POS:DATA?
5.19.4.5 Lower Offset Limit
Sets the limit for the selected offset (refer to
“Offset” above) to the left of the carriers.
Key Path:
Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
FactoryPreset: 0dB,0dB,0dB
Default Terminator: dB
Knob Increment: 1
Step Key
Increment: 5
Range: –200 dB through 200 dB
History: Added with firmware revision A.03.00
Remote Command:
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA <dB>,<dB>,<dB>
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA?
Example: CALC:MCP:OFFS:LIST:LIM:NEG:DATA 50dB,55dB,60dB
CALC:MCP:OFFS:LIST:LIM:NEG:DATA?