ABORt
The ABORt subsystem stops a scan in progress when the scan is enabled via
the interface, and the trigger modes are
TRIGger:SOURce BUS or
TRIGger:SOURce HOLD.
Subsystem Syntax ABORt
Comments ABORt Operation. Resets the trigger system and places triggers in
idle. The scan is stopped without resetting the scan conditions. The
scan conditions stay in the current configuration.
Stopping Scans Enabled from Interface. When a scan is enabled
from the interface, use an interface
CLEAR command or the HP
E1301 front panel "Reset Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS or TRIGger:SOURce HOLD, use ABORt or
the HP E1301 front panel "Reset Instr "or "Clear Instr "keys to stop
the scan.
Stopping Scans Enabled from Front Panel. When a scan is
enabled from the HP E1301 front panel execute
*RST over the
interface or use the front panel "Reset Instr "or "Clear Instr" keys to
stop the scan.
Related Commands. ARM, INITiate:CONTinuous, [ROUTe:]SCAN,
TRIGger
Example Stopping a Scan with ABORt
TRIG:SOUR BUS Bus is trigger source.
INIT:CONT ON Set continuous scanning.
SCAN (@100:115) Sets channel list.
INIT Starts scanning cycle.
.
.
.
ABOR Aborts scan in progress.
56 HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5