HP (Hewlett-Packard) 53A Welder User Manual


 
STATus
The STATus subsystem reports the bit values of the Operation Status
Register. This enables the Status Register to set a bit after a bit is set to 1 by
the Operation Status Register.
Subsystem Syntax STATus
:OPERation
:ENABle <
number>
[:EVENt]?
:OPERation:ENABle STATus:OPERation:ENABle <number> enables the Operation Status
Register to set a bit in the Status Register. For multiplexer modules, when
bit 8 in the Operation Status Register is set to 1, bit 7 in the Status Register
is set to 1.
Parameters
Parameter Name Parameter Type Range of Values
number
Numeric 1 - 32768
Comments Setting Bit 7 of the Status Register: STATus:OPERation:ENABle
256
sets bit 7 of the Status Register to 1 after bit 8 of the Operation
Status Register is set to 1.
Disabling STATus:OPERation: The STATus:PRESet command
disables the
STATus:OPERation:ENABle command.
Related Commands: [ROUTe:]SCAN, *SRE, *STB?, *SPOLL,
STATus:PRESet
Example Enable the Status Register Bit 7
STAT:OPER:ENAB 256 Enables bit 7 of the Status Register.
:OPERation[:EVENt]? STATus:OPERation[:EVENt]? returns the bit value of the Operation
Status Register (only bit 8 is used by the multiplexer modules).
Comments Setting Bit 8 of the Operation Status Register: Bit 8 (Scan
Complete) is set to 1 after a scanning cycle completes. Bit 8 returns
to 0 after sending the
STATus:OPERation[:EVENt]? command. For
correct operation set to 0 by
*CLS before scanning.
Returned Data after sending the STATus:OPERation[:EVENt]?
Command: The command returns +256 if bit 8 of the Operation
Status Register is set to 1. The command returns +0 if bit 8 of the
Operation Status Register is set to 0.
Related Commands: [ROUTe:]SCAN, *SRE, *STB?, *SPOLL
Example Reading the Operation Status Register after a Scanning Cycle
STAT:OPER? Reads the register value
read and print the register value +256 shows bit 8 is set to 1; +0
shows bit 8 is set to 0.
70 HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5