HP (Hewlett-Packard) 53A Welder User Manual


 
SETTling[:TIME] [ROUTe:]SETTling[:TIME] <channel_list> sets the delay between receipt
of a channel closing trigger, and the pulse when the FET channel is closed.
You must specify one and only one channel for each card in the instrument
which has a channel in the channel list, and the
<time> will then apply to all
channels on the card. The range is 1 to 32768
µsec.
Parameters
Parameter Name Parameter Type Range of Values
time
Numeric 1E-6 | 32768E-6 | MIN | MAX
channel_list
Numeric only one channel
Comments Digital Bus Handshaking: This command allows a user-defined
settling time to allow a measurement to be made before the next
channel is closed.
Related Commands: TRIGger:SOURce DBUS, TRIGger:SOURce
IMMediate
Example Setting 16 µsec SETTling TIME
TRIGger:SOUR DBUS Triggering on the digital bus
handshake lines
SETTling:TIME 16E-6 Delay = l6µsec
SETTling[:TIME]? [ROUTe:]SETTling[:TIME]? [MIN | MAX] <channel_list> returns the
current setting for a specified channel.
MIN or MAX returns the minimum
and maximum allowable settings.
Parameters
Parameter Name Parameter Type Range of Values
MIN | MAX Numeric MIN = 1 µsec, MAX = 32768 µsec
channel_list
Numeric only one channel
Comments Returned value is given in exponential notation.
Example Set and Query the SETTling TIME
SETT:TIME 16E-6 Sets delay = 16µsec.
SETT:TIME? Returns + 1.600000E-005.
Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference 69