Interface and Display
AWG5000 Series Quick Start User Manual
29
Run Mode Page
Quick Tip
■ The arbitrary waveform generator supports the following four run modes:
■ Continuous – A continuous waveform is output.
■ Triggered – A waveform is output once when the instrument receives a trigger signal. The instrument will wait for
the next trigger signal after outputting the waveform.
■ Gated – A waveform is output only when a gate signal is asserted. A continuous waveform is output while the gate
signal stays asserted.
■ Sequence – Multiple waveforms can be output in the order specified in the sequence.
1. Select a Run mode.
2. In the Triggered or Gated mode, you
can select the output value while the
instrument is in the waiting-for-trigger
state.
■ First – Sets the output level to the
first value of the waveform
■ Last – Sets the output level to the
last value of the waveform
1