Using the System Menu
68 AWG5000 Series Quick Start User Manual
Setting GPIB/LAN
Quick Tip
■ The following operations cannot be performed through a GPIB or LAN connection:
■ Editing a waveform
■ Changing size or name of waveform
■ Converting the waveform format
■ Importing waveform data from the AWG5000 series and AWG7000 series setup (*.AWG) file
■ Importing a Tektronix DTG5000 series file (*.DTG)
■ Importing a Tektronix AWG400/500/600/700 series SEQ file
■ Exporting a file
Select System > GPIB/LAN Configuration to
open the GPIB/LAN Configuration dialog box.
1. Set the instrument GPIB bus communi-
cation:
■ Talk/Listen – Select this mode to
remotely control the instrument from
an external host computer.
■ Off Bus– Select this mode to discon-
nect the instrument from the GPIB
bus.
2. The instrument supports the following
two LAN connections:
■ VXI-11
■ Raw Socket
1
2