Emerson HASXEDE-IM-EX Welding Consumables User Manual


 
Emerson Process Management GmbH & Co. OHG4-10
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
4.3 Vapor Recovery Application - Special Information
Fig. 4-4: Converter assembly
To replace the converter material
consider all applicable safety instructions
and measures to switch off the analyzer
and connected signals
read the waiting time label on the front
door and wait for the time specied
carefully open the front door (High weight!
Take care not to damage the door!)
The converter is installed behind a metallic
cover, which is xed with a screw nut and a
key (Fig. 4-5). Loosen the screw nut a few
turns (don‘t remove it) and push the cover
upwards until the key is free to remove the
cover (Fig. 4-6).
Now you have access to the inner thermosta-
ted jacket (Fig. 4-6).
Heating elements are inside the
isolating jacket!
Accessible components still may
be hot and harmful to humans!
Fig. 4-5: Converter assembly details
1
2
1: Terminals
2: Metallic cover
4.3.2 Replacement Instructions