Safety Instructions
SAFETY INSTRUCTIONS
Intended Use Statement
X-STREAM XE series gas analyzers are intended to be used as analyzers for industrial purposes.
They must not be used in medical, diagnostic or life support applications.
Using X-STREAM XE analyzers as safety devices is prohibited where redundancy and/or SIL
classication or equivalent is needed.
No independent agency certications or approvals are to be implied as covering such applications!
General Safety Notice / Residual Risk
If this equipment is used in a manner not specied in these instructions, protective systems may
be impaired.
Despite of incoming goods inspections, production control, routine tests and application of state-
of-the-art measuring and test methods, an element of risk remains when operating a gas analyzer!
Even when operated as intended and observing all applicable safety instructions, some residual
risks remain, including, but not limited to, the following:
• An interruption of the protective earth line, e.g. in an extension cable, may result in risk to
the user.
Live parts are accessible when operating the instrument with doors open or covers removed.
• The emission of gases hazardous to health may even be possible when all gas connections
have been correctly made.
Avoid exposure to the dangers of these residual risks by taking particular care when installing,
operating, maintaining and servicing the analyzer.
SPECIAL CONDITIONS FOR SAFE USE
Only specied screws M16x45 ISO 4762
A2-70 as specied in the maintenance
section of this manual shall be used (spare
part # 42716945).
The ame joints correspond to the drawing
No. 4.271-7112/1 and do not comply with
the dimensions mentioned into the Tab. 1
and Tab. 2 of EN 60079-1 ed.2.
The gas path for the sample gas shall be
equipped with additional appropriate ame
arrestors in case of gas pressure above
1100 hPa to 1500 hPa.
Appropriate certied cable glands shall be
used in accordance with IEC/EN 60079-14
Vapor recovery application:
Pressure of gases not to exceed 1100 hPa.
Concentrations of gases must be below
25 % LEL.
Depending on the particular application all
approbriate safety instructions mentioned
in this instruction manual on hand must
be considered!
Take special care of formation of amma-
ble gas at the outlet of breathing and/or
purging devices, if the sample gas con-
centration is above 25% LEL! If need be,
such outlets have to end in a safe area!