A.3 CSA Certicate of Compliance
Certificate:
1714037 (LR 105173)
Master Contract:
185562
Project:
2507282
Date Issued:
March 14, 2012
DQD 507 Rev. 2009-09-01 Page: 2
Gas analyzer, Model: X-STREAM, rated 100-240Vac, 50/60 Hz, 1, 3 – 0,7A, Class I, Pollution Degree II.
• X-STREAM Gas Analyzer Core Compact (XCK) Table Top or Rack Mount with appliance inlet for indoor
use and no display (optional with field wiring terminals for indoor use);
• X-STREAM (X2GK) or X-STREAM Enhanced (XEGK) General Purpose Compact Gas Analyzer:
Table Top or Rack Mount with appliance inlet for indoor use and display (optional with field wiring terminals
for indoor use);
• X-STREAM X100 Compact Gas Analyzer (X100GK): Table Top or Rack Mount with appliance inlet for
indoor use and display (optional with field wiring terminals for indoor use);
Gas analyzer, Model: X-STREAM, rated 100-240Vac, 50/60 Hz, 5.5 - 3A, Class I, Pollution Degree II.
• X-STREAM (XXF) or X-STREAM Enhanced Field Housing Gas Analyzer (XDF): Wall mounting with
field wiring terminals, for outdoor use type 4 & IP66 and display;
Gas analyzer, Model: X-STREAM, rated 24Vdc, 2.5A, Class I, Pollution Degree II.
• X-STREAM (X2GC or X2GK)or X-STREAM Enhanced (XEGC or XEGK)General Purpose Compact
Gas Analyzer Table Top or Rack Mount with 24Vdc in connector and display;
• X-STREAM Compact Gas Analyzer Core (XCC or XCK): Table Top or Rack Mount with 24Vdc in
Conditions of Acceptability
- For the X-STREAM Models X2GP, XCA and XEGP and the AC powered versions of X2GK, X2CK, XEGK
and X100GK the equipment is supplied with an approved power supply cord set or power supply cord with plug
that is acceptable to the authorities in the country where the equipment is to be used. Units supplied without a
power cord and that are not permanently connected are considered as component. Component-type units must
be provided with a Fire, Mechanical and Electrical enclosure and must be re-evaluated by CSA.
- The plug/connector is used as the disconnected device. The switch for X2GP/XCA/XEGP/XCK/X2GK/
XEGK/X100GK is not considered the disconnect device. All units must be provided with a disconnect device.
CLASS 2258-02 PROCESS CONTROL EQUIPMENT – For Hazardous Locations
CLASS 2258-82 PROCESS CONTROL EQUIPMENT – For Hazardous Locations – Certified to U.S.
Standards.
X-Stream FD (XFD): Flameproof for Hazardous Locations
Class I, Zone 1, Ex d IIB+H2, T3 and/or Class I, Division 2, Groups B, C, and D, T3
Class I, Zone 1, AEx d IIB+H2, T3 and/or Class I, Division 2, Groups B, C, and D, T3
Gas analyzer, Model: X-Stream, rated 100-240Vac, 50/60 Hz, 2–1 A. Class I, Pollution Degree II; Type 4 &
IP66
Ambient Temperature Range: -30°C to +50°C Maximum internal case pressure = 110kpa