4 Service and Maintenance
EXPLOSION HAZARD BY FLAMMMABLE GASES
Leaks may cause explosion when measuring ammable gases!
When measuring ammable gases it is recommended to perform a leak test
on all gas paths, connections and components before startup or applying
power. Leak tests should be carried out on a 2 month‘s regular basis and
after repair/maintenance.
See the analyzer instruction manual for instructions on how to carry out
leak tests.
HAZARDS BY HAZARDOUS GASES
When measuring ammableand/or toxic gases it is recommended to purge
the system with air or an inert gas, e.g. nitrogen, prior to opening the gas
paths.
Violation may cause an explosion and/or personal injury or death!
EXPLOSION HAZARD BY UNAUTHORIZED MODIFICATION
Modications affecting the integrity of type of protection (e.g. afxing
addtional threads, replacing the ame arrestors by other model) are NOT
PERMITTED!
Violation may cause an explosion and/or personal injury or death!