Emerson Process Management GmbH & Co. OHG 4-17
4
Maintenance
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
Zero Calibration..
Span Calibration..
Advanced Calibration..
Apply Gas..
Lock Menus!
Acknowledgements
Pump 1: Off
Pump 2: Off
Ranges..
4.4 Manual Calibration
When nished, press left several times to re-
turn to SELECT COMPONENT (multi channel
analyzer only), to perform a span calibration
for another channel,
or
press home to return to the MEASUREMENT
SCREEN, to nish with manual calibration
procedures.
4.4.2 Manual Span Calibration
Multi-channel unit:
Select the channel to be calibrated in
SELECT COMPONENT.
Before selecting any further line
make sure the required calibra-
tion gas is applied and owing!
SPAN CALIBRATION offers the similiar opti-
ons as ZERO CALIBRATION.
Select "Span Calibration.."
Span Calibration
Cancel!
Start!
Span Gas 0.000 ppm
Concentration 0.000 ppm
Flow 0.00 l/min
Status..
Results..
Restore..