Safety Instructions
ELECTRICAL SHOCK HAZARD
Installation, and connecting mains and signal cables are subject to qualied
personnel only, taking into account all applicable standards and legislative
requirements!
Failure to follow may cause warranty invalidation, property damage and/or
personal injury or death! Connecting mains and signal cables to internal
srew terminals requires working at open housing near life parts!
Installation of this instrument is subject to qualied personnel only, familiar
with the resulting potential risks!
The gas analyzers do not provide a mains power switch and are operable
when connected to power.
The gas analyzers do not provide a mains switch! A mains switch or circuit
breaker (to comply with IEC 60947-1 /-3) has to be provided in the building
installation. This switch has to be installed near by analyzer, must be
easily operator accessible and has to be assigned as disconnector for the
analyzer.
EXPLOSION and ELECTRICAL SHOCK HAZARD BY INSTALLATION
These instruments provide a protective earth terminal. To prevent electrical
shock and explosion hazards, the instrument must be connected to a
protective earth.
Therefore the instrument has to be connected to mains by using a three
wire mains cable with earth conductor!
Any interruption of the earth connector inside or outside the instrument
or disconnecting the earth terminal may cause potential electrical shock
hazard!
Intended interruption of protective earth connections is not permitted!