4.1 Verications and Tests
4.1 Verications and Tests
4.1.1 Routine Tests
The following tests shall be performed on a
regular basis:
EXPLOSION HAZARD BY DAMAGED ANALYZER
Do not continue to operate a damaged analyzer!
Call for service!
Violation may cause an explosion and/or personal injury or death!
4.1.1.1 Visual Inspection
The instrument shall be visually checked for
damages to the enclosure, ange, threads
and external analyzer components.
Especially take care that there are no scrat-
ches on the anges surfaces!
Contents
4.1 Verications and Tests ........................................4-3
4.2 Replacement of Parts.........................................4-6
4.3 Vapor Recovery Application - Special Information ...................4-8
4.4 Perform a calibration ........................................4-15
In the case of repairs affecting the type of
protection, the parts which have been re-
paired should be subjected to new routine
verications and tests. These tests need not
necessarily be made by the manufacturer.
4.1.1.2 Detailed Inspection
Check cable inlets (glands or conduits): If
the cable moves, e. g. thighten the cap nut.
If tightening is ineffective, replace the cable
gland or conduit.
Also consider the original manufacturer‘s
instructions for maintenance!