2.4 Installation - Electrical
EXPLOSION HAZARD WHEN INSTALLING CABLE GLANDS
When using the ADL type cable glands: Take care to use the correct cable
gland internal elastomeric ring with an internal diameter supporting the cable
to be used! The supported cable diameters are noted on the sealing ring.
Use only cables as specied in the cable glands installation instruction,
and carefully follow the installation instructions!
EXPLOSION HAZARD BY INTERNAL BATTERY
This analyzer contains an internal battery!
Do NOT OPEN the analyzer enclosure if explosive atmosphere may be
present!
Disregarding may cause explosion even if the analyzer is de-energized!
The next page shows
• the English installation instructions for type
AD-U adapters, reprinted with the permis-
sion of
Redapt Ltd
Note!
See the documentation delivered together
with your analyzer for an original version of
the cable gland or adapter installation instruc-
tions, whatever is applicable !