Safety Instructions
HOW TO STAY IN COMPLIANCE WITH THE
EUROPEAN DIRECTIVE 94/9/EC ("ATEX") WHEN PERFORMING GAS
ANALYSIS WITHIN A FLAMEPROOF ENCLOSURE.
Special conditions apply to using a ameproof enclosure analyzer under the scope of the
"European Directive for Equipment used in Explosive Atmosphere" (Directive 94/9/EC;
ATEX). To stay compliant with the directive, consider the following clarication sheet re-
leased by the European ATEX Notied Body Group (see next page):
HEAVY INSTRUMENT HAZARD
X-STREAM XEFD analyzers, to which this manual relates, intended
to be wall mounted and/or outdoor installed, weigh up to approx. 63 kg
(139 lbs), depending on included options!
Use two people and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts specied to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid
and stable to hold the units!
HIGH TEMPERATURES HAZARD
While working at internal components hot surfaces may be accessible, even
after the instrument has been disconnected from power!
EXPLOSION HAZARD: VAPOR RECOVERY APPLICATION
Consider the waiting time statement on the front door label before opening!
Disregarding the waiting time statement may result in explosion!
Temperatures of components for this application installed into the analyzer
exceed the analyzer's temperature classication for hazardous areas!