Emerson HASXEDE-IM-EX Welding Consumables User Manual


 
Emerson Process Management GmbH & Co. OHG 1-1
1
Techn. Description
X-STREAM XEFD
Instruction Manual
HASXEDE-IM-EX
03/2012
Chapter 1
Technical Description
1.1 Overview
The new X-STREAM ameproof gas analy-
zers are designed to be used in hazardous
areas. The flameproof enclosure can be
installed at Zone 1 and Division 2 locations
without the need of any additional protective
measures, e.g. purge gas supply.
1.2 Design Features
Packaged into a cast aluminum enclosure,
the X-STREAM XEFD gas analyzer provides
all the measurement options available for ge-
neral purpose instruments, but for installation
at locations, where explosive gas atmosphere
might be present frequently (Zone 1) or occa-
sionally (Zone 2, Division 2).
The basic concept used to protect the sur-
rounding atmosphere from being ignited, if an
internal failure results in high temperatures,
ames or even an explosion, is to keep the
explosion inside the enclosure and quench all
ames possibly passing through the ange.
To provide adequate explosion protection the
X-STREAM ameproof analyzer feature:
•
a cast aluminum enclosure, designed to
• withstand an internal explosion,
• quench ames resulting from an internal
explosion (thus preventing a surroun-
ding explosive atmoshere from being
ignited).
• ame arrestors avoiding ame transmissi-
on from the gas paths into the surrounding
atmosphere.
• approved cable glands (option: conduits),
protecting the cable inlets and outlets.
Fig. 1-1: Frontal View
1: Enclosure base
2: Screws
3: Enclosure cover
4: Flange
5: Eyebolts for lifting
6: Hinges
1
2
3
5
4