Tektronix TDS510A Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–99
DATa:DESTination
Sets or queries the reference memory location for storing waveform data that is
transferred into the digitizing oscilloscope by the CURVe command. This
command is identical to the DATa:TARget command.
Waveform
DATa:DESTination REF<x>
DATa:DESTination?
DATa
:
DESTination
?
<Space> <x>
REF
REF<x> (REF1, REF2, REF3 or REF4) is the reference memory location where the
waveform will be stored.
This means that you cannot have a channel as a destination. This is true even for
extended-acquisition-length mode.
DATA:DESTINATION REF3
stores incoming waveform data in reference memory 3.
DATA:DESTINATION?
might return REF2 as the reference memory location that is currently selected.
DATa:ENCdg
Sets or queries the format of the waveform data. This command is equivalent to
setting WFMPre:ENCdg, WFMPre:BN_Fmt, and WFMPre:BYT_Or as shown
in Table 2–30. Setting the DATa:ENCdg value causes the corresponding
WFMPre values to be updated and vice versa.
Waveform
WFMPre:ENCdg, WFMPre:BN.FMT, WFMPre:BYT_Or
DATa:ENCdg { ASCIi | RIBinary | RPBinary | SRIbinary | SRPbinary}
DATa:ENCdg?
Group
Syntax
Arguments
Examples
Group
Related Commands
Syntax