Command Descriptions
2–52
TDS Family Oscilloscope Programmer Manual
ACQuire:NUMEnv
Sets the number of waveform acquisitions that make up an envelope waveform.
This is equivalent to setting the Envelope count in the Acquisition Mode side
menu.
Acquisition
ACQuire:MODe
ACQuire:NUMEnv { <NR1> | INFInite }
ACQuire:NUMEnv?
ACQuire
:
<NR1>
?
NUMEnv
<Space>
INFInite
<NR1> 0 is the number of waveform acquisitions, from 1 to 2000. The
envelope will restart after the specified number of envelopes have been acquired
or when the ACQuire:STATE RUN command is sent.
INFInite or <NR1> = 0 specifies continuous enveloping.
NOTE. If you set the acquisition system to single sequence, envelope mode, and
set the number of envelopes to infinity, the digitizing oscilloscope will envelope a
maximum of 2001 acquisitions.
ACQUIRE:NUMENV 10
specifies that an enveloped waveform will show the result of combining 10
separately acquired waveforms.
ACQUIRE:NUMENV?
might return 0, indicating that acquisitions are acquired infinitely for enveloped
waveforms.
Group
Related Commands
Syntax
Arguments
Examples