Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–165
the 1–2–5 sequence (1–2.5–5 on the TDS 620B, 640B, & 644B), it is automati-
cally set to the closest valid value.
HORIZONTAL:DELAY:SCALE?
might return 1.0E-3, indicating that the delay time is 1 ms per division.
HORizontal:DELay:SECdiv
This command is identical to the HORizontal:DELay:SCAle command. It is
provided to maintain program compatibility with some older models of
Tektronix digitizing oscilloscopes.
Horizontal
HORizontal:DELay:SECdiv <NR3>
HORizontal:DELay:SECdiv?
:
SECdiv
:
DELay
HORizontal
?
<Space> <NR3>
HORizontal:DELay:TIMe
Sets or queries the delay time to wait after the main trigger before the delayed
time base begins. This is equivalent to setting Delayed Runs After Main in the
Time Base side menu of the Horizontal menu.
Horizontal
HORizontal:DELay:MODe
HORizontal:DELay:TIMe <NR3>
:
DELay
:
HORizontal
TIMe <NR3><Space>
<NR3> is the time, in seconds, between the main trigger and the delayed trigger.
The range on the TDS 510A, 500B, 600B, & 700A is from 16 ns to 250 seconds
with a resolution of 4 ns. The range on the TDS 400A is from 10 ns to 20
seconds with a resolution of 10 ns.
Group
Syntax
Group
Related Commands
Syntax
Arguments