Command Descriptions
2–324
TDS Family Oscilloscope Programmer Manual
this argument, it will convert it to LINE. The TDS 400A will not output this
argument in response to a query.
LINE specifies a delay by a number of video lines.
TRIGGER:MAIN:VIDEO:BY TIME
specifies a delay by time.
TRIGger:MAIn:VIDeo:FIELD
TDS 400A Option 05 Only
Sets or queries the field the video trigger acts on. This is equivalent to using the
Video Scan side menu when Class is NOT set to Custom.
Trigger
TRIGger:MAIn:VIDeo:FIELD {ĂODD | EVEN | ALL | FIELD1 | FIELD2 |
FIELDEitherĂ}
TRIGger:MAIn:VIDeo:FIELD?
FIELDEither
TRIGger :
:
VIDeo
MAIn
FIELD
:
<Space>
FIELD1
?
FIELD2
ALL
EVEN
ODD
ODD specifies interlaced video field 1.
EVEN specifies interlaced video field 2.
ALL specifies alternating both video field 1 and video field 2.
FIELD1 specifies interlaced video field 1. For the TDS 400A, this argument is
available only for backward compatibility. If the TDS 400A receives this
argument, it will convert it to ODD. The TDS 400A will not output this
argument in response to a query.
FIELD2 specifies interlaced video field 2. For the TDS 400A, this argument is
available only for backward compatibility. If the TDS 400A receives this
Examples
Group
Syntax
Arguments