Tektronix TDS510A Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–219
MEASUrement:MEAS<x>? (Query Only)
Returns all measurement parameters for the displayed measurement specified by
<x>.
Measurement
MEASUrement:MEAS<x>?
:MEASUrement
MEAS
<x> ?
MEASUREMENT:MEAS3?
might return: :MEASUREMENT:MEAS3:STATE 0;TYPE PERIOD;
UNITS "s";SOURCE1 CH1;SOURCE2 CH2;DELAY:EDGE1 RISE;
EDGE2 RISE;DIRECTION FORWARDS.
MEASUrement:MEAS<x>:COUNt? (Query Only)
TDS 500C & 700C Only
Returns the number of values accumulated for this measurement since the last
statistical reset. Some values may have been ignored because they generated an
error.
Measurement
MEASUrement:MEAS<x>:COUNt?
:MEASUrement
MEAS
<x> ?
:
COUNt
<NR3>
MEASUREMENT:MEAS1:COUNT?
might return: :MEASU:MEAS1:COUN 3.247000000E+3
MEASUrement:MEAS<x>:DELay? (Query Only)
Returns the delay measurement parameters for the measurement specified by
<x>.
Group
Syntax
Examples
Group
Syntax
Returns
Examples