Tektronix TDS510A Caulking Gun User Manual


 
Command Descriptions
2–168
TDS Family Oscilloscope Programmer Manual
HORizontal:EXTDACQ
TDS 500C & 700C Only
Enable or disable extended-acquisition-length mode. The InstaVu display mode
must be off in order to turn the extended-acquisition-length mode on.
Horizontal
HORizontal:EXTDACQ { <NR1> | OFF | ON }
HORizontal:EXTDACQ?
:HORizontal
?
<Space>
<NR1>
ON
OFF
EXTDACQ
<NR1> indicates OFF if it’s a 0 or ON if it’s a 1 (or any other nonzero value).
ON means turn on.
OFF means turn off.
The query form only returns 0 or 1. The default is 0.
HORIZONTAL:EXTDACQ ON would enable extended-acquisition-length mode
(if InstaVu was not on).
HORizontal:FASTframe:COUNt
TDS 500C & 700A Only
Sets or queries FastFrame frame count. This is equivalent to setting FastFrame
Setup in the Horizontal menu and the Frame Count menu item in the side
menu. FastFrame, also known as memory segmentation, captures a series of
triggered acquisitions with minimal, intervening, time.
Horizontal
HORizontal:FASTframe:COUNt <NR1>
Group
Syntax
Arguments
Returns
Examples
Group
Syntax