Command Descriptions
2–312
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:RUNT:WIDth
TDS 500C, 600B, & 700C Only
Sets or queries the minimum width for a valid main pulse runt trigger. This
command is equivalent to entering a value in the Trigger menu’s Wider Than
side menu.
Trigger
TRIGger:MAIn:PULse:RUNT:WIDth <NR3>
TRIGger:MAIn:PULse:RUNT:WIDth?
TRIGger
:
:
PULse
MAIn
RUNT
:
:
?
<Space>
<NR3>
WIDth
<NR3> is the minimum width in seconds.
TRIGGER:MAIN:PULSE:RUNT:WIDTH 15E-6
sets the minimum width of the pulse runt trigger to 15 ms.
TRIGger:MAIn:PULse:SLEWRate:DELTATime
TDS 500C, 600B, & 700C Only
Sets or queries the delta time used in calculating the slew rate trigger. This is
equivalent to selecting Trigger When from the main Trigger’s Slew Rate menu
and Delta Time in the resulting side menu.
Trigger
TRIGger:MAIn:PULse:SLEWRate:DELTATime <NR3>
TRIGger:MAIn:PULse:SLEWRate:DELTATime?
Group
Syntax
Arguments
Examples
Group
Syntax