FACtory (No Query Form)
Resets the digitizing oscilloscope to its factory default settings. This command is
equivalent to selecting Recall Factory Setup in the Waveform Save/Recall
menu.
Miscellaneous
*PSC, *RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
FACtory
FACtory
Setting the digitizing oscilloscope to factory default includes:
H Clears the Event Status Enable Register.
H Clears the Service Request Enable Register.
H Sets the Device Event Status Enable Register to 255.
H Sets the Power On Status Clear Flag to TRUE
H Purges all defined aliases.
H Enables all Command Headers (HEADer ON).
H Sets the macro defined by *DDT to a “zero-length field.”
H Clears the pending operation flag and associated operations.
The FACtory command does not alter the following:
H The state of the GPIB (IEEE Std 488.1-1987) interface.
H The selected GPIB address.
H Calibration data that affects device specifications.
H Protected user data.
H Stored settings.
H The current password (if implemented).