Command Descriptions
2–314
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:SLEWRate:SLEWRate? (Query Only)
TDS 500C, 600B, & 700C Only
Returns the slew rate value. This is the
((Upper Threshold – Lower Threshold) / Delta Time)
The value is limited to the three most significant digits.
Trigger
TRIGger:MAIn:PULse:SLEWRate:SLEwrate?
TRIGger :
:
PULse
MAIn
:
?SLEWRate
: SLEWRate
TRIGGER:MAIN:PULSE:SLEWRATE:SLEWRATE?
returns the slew rate. It is given as an appropriate amount of volts per second.
For example, 1.2E+6 would represent a slew rate of 1.2 megavolts/second.
TRIGger:MAIn:PULse:SLEWRate:THReshold:BOTh
TDS 500, 600B, & 700C Only
Sets the upper and lower slew rate trigger thresholds. This is equivalent to
selecting Thresholds from the main Trigger’s Slew Rate menu and entering a
value in the resulting side menu’s High Threshold or Low Threshold items.
Trigger
TRIGger:MAIn:PULse:SLEWRate:THReshold:BOTh { TTL | ECL }
THReshold
TRIGger :
:
PULse
MAIn
:
:
BOTh
SLEWRate
ECL
TTL
TTL sets the upper threshold to 1.8 V and the lower threshold to 0.8 V.
ECL sets the upper threshold to –1.1 V and the lower threshold to –1.5 V.
Group
Syntax
Examples
Group
Syntax
Arguments