Sensor I® Expert User Manual
Dynamic Audit
1. Press and release the Reset button to power
up the wrench (if the wrench is left idle for
one minute the wrench will turn off to save
battery power).
2. Move the Mode Selection switch to Dynamic
Audit.
3. Press the Units key to select the desired
Torque units.
4. Attach the angle reference arm (EA-models
only) to the wrench, and place the magnet
on a magnetic surface.
5. Press and hold in the Reset button. The cap-
ture angle will be displayed.
6. With the Reset button held in, use the
parameter adjustment knob to adjust the
capture angle to the desired value (see the
application guide for details). Release the
Reset button.
7. Press the Display key until "Pk" is displayed
on the lower left-hand side of the screen.