Section 7: Native Mode Commands In-Sight
®
1700 Series Wafer Reader
66
7.2.3 Put Watch
Put Watch(<Column><Row><Input>)
Returns the data contained in the specified cell each time the cell is updated.
NOTE The PortNum command must be executed to specify a port where the information will be sent by
the Watch command.
Inputs: Column, Row, Input
Output: Status Code
The status codes are:
STATUS CODE DESCRIPTION
1 The command was executed successfully.
0 Unrecognized command.
-2 The command could not be executed.
Sample Input and Output
PUT WATCH A000 1
1
7.3 Inputs and Outputs
The table below list the possible values of the settings that can be input and possible
values of the settings and results data returned by the Evaluate Native Mode commands.
These output values are enclosed in Extensible Markup Language (XML) tags (XML is a
web document authoring language which allows users to create their own customized
tags. For more information on the XML format, go to http://www.w3.org/XML).
SETTING DESCRIPTION
<Accept>
The minimum response threshold for the raw reading score.
40 – 100 (default = 50)
<AcqEnable>
The status of the Disable Acquire setting.
0 = Disable
1 = Enable
<AcquireFlag>
Determines whether an image will be acquired when a read is performed.
0 = A read will be performed on the last image acquired.
1 = A read will be performed after a new image is acquired.
<AvgScore>
The average raw response score of all reads that passed for a Config.
0 – 100
<BestConfig>
The index of the Config which returns the highest score.
0 – 9
<Bright>
Determines whether Bright Field Light Mode settings will be tried during tuning.
0 = Disable Bright Field tuning
1 = Enable Bright Field tuning