252 Chapter4
Instrument Functions: P - Z
System
Instrument Functions: P - Z
Default: Mode
Remote Command:
See “Preset Type” on page 250.
Example: SYST:PRES:TYPE FACT defines the type of preset as the factory preset.
With factory preset selected, use SYST:PRES to do a factory preset.
4.8.2.3 Save User Preset
Saves the current state of the analyzer into the
User Preset state registor for recall when the
instrument Preset Type is set to User and you perform a preset operation. After you save a
statehere,youmustgotothe
Preset Type key and select User in order to have this state used
as the preset state. Refer to
“Preset” on page 221 key description for the default
factory-configuration settings.
Key Path:
System, Power On/Preset
Preset
,ifthepresettypeissettoUser.See“Save User Preset” on page 225
Annunciation/
Annotation: None
Dependencies/
Couplings: None
SCPI Status Bits/
OPC Dependencies: None
Remote Command:
:SYSTem:PRESet[:USER]:SAVE
Example: SYST:PRES:SAVE
Use SYST:PRES:TYPE USER to set factory preset type to “User.”
Then use SYST:PRES to do the preset.
4.8.3 Time/Date
Displays the Time/Date function menu keys used to set and display the real-time clock.
Key Path:
System
Remote Command:
There is no remote command for this key.