Agilent Technologies E4440-90189 Saw User Manual


 
Chapter 5 415
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
Remote Command:
This parameter can send up to 10 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted. For example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values remain
as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:STATe OFF|ON|0|1
The query for this parameter will always return 20 values.
[:SENSe]:HARMonics:RANGe[:LIST]:STATe?
Example: HARM:RANG:LIST:STAT ON,OFF,ON,ON
5.16.6.2 Center Freq
Used to set the center frequency of the analyzer.
Key Path:
Meas Setup, Range Table
Factory Preset: See Table 5-1 on page 414.
Knob Increment: 1%
Step Key
Increment: 1, 1.5, 2, 3, 5, 7.5, 10, ...
Range: Frequency range of your analyzer.
History: Added with firmware revision A.03.00
Remote Command:
This parameter can send up to 10 values. The location in the list sent corresponds to the
range the value is associated. Missing values are not permitted, for example, if you want to
change values 2 and 6, you must send all values up to 6. Subsequent values will remain as
they were.
[:SENSe]:HARMonics:RANGe[:LIST]:FREQuency <integer>
The query for this parameter will always return 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:FREQuency?
Example: HARM:RANG:LIST FREQ 300
5.16.6.3 Span
Span is used to set the span of the analyzer.
Key Path:
Meas Setup, Range Table
Factory Preset: See Table 5-1 on page 414.