Chapter 5 409
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
5.16 Meas Setup (Harmonic Distortion)
When the harmonic distortion measurement has been selected in the Measure key menu of
the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
Key Path: Front-Panel key
Saved State: Not saved.
5.16.1 Avg Number
Press Avg Number (On) to specify the number of measurement averages used when
calculating the measurement result. The average will be displayed at the end of each
sweep.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 averages / Off
Range: 1 to 1000
Remote Command:
[:SENSe]:HARMonics:AVERage:COUNt <integer>
[:SENSe]:HARMonics:AVERage:COUNt?
[:SENSe]:HARMonics:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HARMonics:AVERage[:STATe]?
Example: HARM:AVER:COUN 100
HARM:AVER ON
5.16.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function to
either Exp or Repeat. This determines the averaging action after the specified number of
measurements (average count) is reached.
EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive data
acquisition after the average count is reached is exponentially weighted and combined
with the existing average. Exponential averaging weights new data more than old data,
which facilitates tracking of slow-changing signals. The average will be displayed at the
end of each sweep.