Chapter 5 491
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]?
5.27.3.5 Video BW
Video BW is used to set the video bandwidth of the analyzer. When Auto is selected the
analyzer determines the optimum setting, while
Man enables you to determine the setting.
Key Path:
Meas Setup, Range Table
Factory Preset: See “Range Table” on page 486.
Knob Increment: Steps through the available video bandwidth filters.
Step Key
Increment: Steps through the available video bandwidth filters.
Range: Video bandwidth range of your analyzer.
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo:AUTO OFF|ON|0|1
selects the mode.
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo <integer>
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth|BANDwidth:VIDeo?
5.27.3.6 Sweep Time
Sweep Time is used to set the sweep time of the analyzer. When Auto is selected the analyzer
determines the optimum setting, while
Man enables you to determine the setting.
Key Path:
Meas Setup, Range Table
Factory Preset: See “Range Table” on page 486.
Knob Increment: 1%
Step Key
Increment: 1, 1.5, 2, 3, 5, 7.5, 10,...
Range: Sweep time range of your analyzer.