Agilent Technologies E4440-90189 Saw User Manual


 
334 Chapter5
One-Button Measurement Functions
Mode Setup (Spectrum Analysis Mode)
One-Button Measurement Functions
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect]
See“RadioStd”onpage333.
Example: RAD:STAN NONE
5.3.1.2 IS-95A
Sets the specific parameters for the selected measurement (located under the
“MEASURE
(Spectrum Analysis Mode)”
key description) appropriate for industry standard IS-95A. The
Spectrum Emissions Mask, Harmonic Distortion, and Spurious Emissions measurements
are not available when this radio standard is selected.
Key Type: 1 of N menu
Key Path:
Mode Setup, Radio Std
History: Added with PSA firmware revision A.02.00
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect]
See“RadioStd”onpage333.
Example: RAD:STAN IS95
5.3.1.3 J-STD-008
Sets the specific parameters for the selected measurement (located under the
“MEASURE
(Spectrum Analysis Mode)”
key description) appropriate for industry standard J-STD-008.
The Spectrum Emissions Mask, Harmonic Distortion, MCP, and Spurious Emissions
measurements are not available when this radio standard is selected.
Key Type: 1 of N menu
Key Path:
Mode Setup, Radio Std
History: Added with PSA firmware revision A.02.00
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect]
See“RadioStd”onpage333.
Example: RAD:STAN JSTD
5.3.1.4 IS-95C
Sets the specific parameters for the selected measurement (located under the
“MEASURE
(Spectrum Analysis Mode)”
key description) appropriate for industry standard IS-95C. The
Spectrum Emissions Mask, Harmonic Distortion, and Spurious Emissions measurements
are not available when this radio standard is selected.
Key Type: 1 of N menu