258 Chapter4
Instrument Functions: P - Z
System
Instrument Functions: P - Z
Factory Preset: On
Factory
Default: On
Remote Command:
:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Example: CAL:FREQ OFF
4.8.4.4 Align Subsys
Displays keys that enable you to activate a partial alignment.
Key Path:
System, Alignments
Remote Command:
There is no remote command for this key.
4.8.4.4.1 Align RF
Initiates an alignment on the RF assembly.
Key Path:
System, Alignments, Align Subsys
Annunciation/
Annotation: Will display a series of pop-up message boxes, indicating an alignment is
in progress.
If a signal is present which interfere with the alignment a message Align
RF 50 MHz signal detected. will be displayed.
SCPI Status Bits/
OPC Dependencies: Status questionable calibration. Bits 3, 11, or 12 will be set if
Align RF
fails
Remote Command:
:CALibration:RF
:CALibration:RF?
Remote Command Notes: The query performs the alignment and returns a zero if the
alignment is successful.
Example: CAL:RF?
4.8.4.4.2 Align IF