Agilent Technologies E4440-90189 Saw User Manual


 
360 Chapter5
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
[:SENSe]:ACPower:TYPE?
Example: ACP:TYPE PSDR
ACP:TYPE?
5.6.6 Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, in which it is
free from errors due to input compression. This setting will not necessarily give the
optimum dynamic range, nor the optimum accuracy. No single setting can optimize both,
and the optimum setting often depends on the signal characteristics.
NOTE There will always be a minimum of 2 dB of attenuation set to protect the
analyzer input.
Key Path:
Meas Setup
State Saved: Not saved.
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example: POW:RANG:AUTO ONCE
5.6.7 Method
Enables you to set the measurement method to either the integration bandwidth method
(IBW) or the resolution bandwidth method (RBW). The resolution bandwidth method is
most useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth
method is preferred for other signals.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: IBW
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:METHod IBW|RBW
[:SENSe]:ACPower:METHod?
Example: ACP:METH RBW
ACP:METH?