372 Chapter5
One-Button Measurement Functions
Meas Setup (Burst Power)
One-Button Measurement Functions
5.8.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function to
either Exp or Repeat. This determines the averaging action after the specified number of
measurements (average count) is reached.
• EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive data
acquisition after the average count is reached is exponentially weighted and combined
with the existing average. Exponential averaging weights new data more than old data,
which facilitates tracking of slow-changing signals. The average will be displayed at the
end of each sweep.
• Repeat—When you set Avg Mode to Repeat, after reaching the average count, all previous
result data is cleared and the average count is set back to 1.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: EXPonential
Range: EXPonential|REPeat
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:BPOWer:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:BPOWer:AVERage:TCONrol?
Example: BPOW:AVG:TCON EXP
BPOW:AVG:TCON?
5.8.3 Average Type
Allowsyoutospecifythetypeofresultaveragingtobeperformed.
• Log — Selects averaging that sums the trace data and divides by the number of data
points.
• RMS — Selects averaging that converts trace data from dB to power units, then averages
the power trace data. This selection requires more time to perform.
Key Path:
Meas Setup
Factory Preset: RMS
Range: RMS or Log
History: Added with firmware revision A.02.00
Remote Command:
[:SENSe]:BPOWer:AVERage:TYPe LPOWer|POWer