306 Chapter5
One-Button Measurement Functions
MEASURE (Spectrum Analysis Mode)
One-Button Measurement Functions
Remote Command:
Measurement Results Available
Condition Command Results Returned
:CONFigure:ACPower
:INITiate:ACPower
Not Applicable
RadioStd=Noneand
only offset A is on
:FETCh:ACPower[n]?
:MEASure:ACPower[n]?
:READ:ACPower[n]?
Returns 3 comma-separated values that
correspond to:
main channel power,
lower-adjacent channel power (dBc), and
upper-adjacent channel power (dBc).
Radio Std None
or
RadioStd=Noneand
more than one
offset is on
Meas Type = Total Pwr
Ref
Returns 28 comma-separated values that
correspond in the following order:
1. Main Channel- relative power (dB)
2. Main Channel- absolute power (dBm)
3. Main Channel- relative power (dB)
4. Main Channel- absolute power (dBm)
5. Negative Offset Frequency (1)
- relative power (dB)
6. Negative Offset Frequency (1)
- absolute power (dBm)
7. Positive Offset Frequency (1)
- relative power (dB)
8. Positive Offset Frequency (1)
- absolute power (dBm)
...
25. Negative Offset Frequency (6)
- relative power (dB)
26. Negative Offset Frequency (6)
- absolute power (dBm)
27. Positive Offset Frequency (6)
- relative power (dB)
28. Positive Offset Frequency (6)
- absolute power (dBm)
Main channel relative power is relative to
the main channel absolute power, and
therefore is always equal to 0.00 dB.